American Cleanroom Systems
Request A Quote

The Top 5 Mistakes Made In Particle Testing Cleanrooms

Bench top cleanroom particle counter, metal case, LCD screen
Figure 1. Particle Counter

Cleanroom particle testing standards are set by ISO-14644-1 and 2. Make sure you follow these standards when particle testing your cleanroom. The IEST (the Institute of Environmental Sciences and Technology) works with ISO to write the testing standards. Regular correctly executed particle tests are necessary to ensure the proper operation of your cleanroom. This blog covers commonly made mistakes in cleanroom particle test.

ISO 14644-1 Cleanroom Standards | Cleanroom Classifications
Class Maximum  Particles/m³ FED STD 209E equivalent
>0.1 um >0.2 um >0.3 um >0.5 um >1 um >5 um
ISO 1 10 2
ISO 2 100 24 10 4
ISO 3 1,000 237 102 35 8 Class 1
ISO 4 10,000 2,370 1,020 352 83 Class 10
ISO 5 100,000 23,700 10,200 3,520 832  29 Class 100
ISO 6 1,000,000 237,000 102,000 35,200 8,320 293 Class 1,000
ISO 7 352,000 83,200 2,930 Class 10,000
ISO 8 3,520,000 832,000 29,300 Class 100,000
ISO 9 35,200,000 8,320,000 293,000 Room Air
Table 1. ISO cleanroom classifications

Mistake #1:  Not taking enough sample locations in your cleanroom

ISO-14644-1 2015 significantly increased the number of sample locations required to test a cleanroom.   Be sure you are taking the correct number of sample locations.

Area of room m² Area of room f ² ISO-14644-1: 1999 ISO-14644-1 2015
52 560 8 10
56 603 8 11
64 689 8 13
68 732 9 13
72 775 9 14
76 818 9 15
104 1119 11 16
108 1163 11 17
116 1249 11 18
148 1593 13 19
156 1679 13 20
192 2067 14 21
232 2497 16 22
276 2971 17 23
352 3789 19 24
436 4693 21 25
636 6845 24 26
1000 10764 32 27
>1000 >10764 n/a See Formula
Table 2. IS-14644 number of cleanroom test locations

For rooms large than 1000 m2 use the following formula:

NL = 27 x (Area m2 / 1000)

Mistake #2: Using the wrong type of particle counter to test your cleanroom

Handheld particle counters should only be used to look for particle sources. They do not test a large enough volume of air to accurately measure particle density in a cleanroom.

Hand held cleanroom particle counter
Figure 2. Hand held particle counter

Bench top particle counters should be used to measure particle density in cleanrooms. Benchtop particle counters typically measure. Lower flow rate (0.1 cfm) cleanroom particle counters can be used for lower class ISO-7 cleanrooms and ISO-8 cleanrooms. Higher flow rate  (1.0 cfm) cleanroom particle counters are required for higher class ISO-5 cleanrooms and ISO-6 cleanrooms.

Bench top cleanroom particle counter, large LCD screen
Figure 3. Benchtop particle counter

Mistake #3:  Not getting your particle counter calibrated annually

Particle counters should be calibrated annually by manufacturer or licensed 3rd party service with  NIST traceable standard. Properly calibrated equipment is a requirement to obtain an accurate particle count measurement.

Pharmaceutical cleanroom interior, stainless steel mixers, personnel in bunny suits
Figure 4. Pharmaceutical cleanroom in operation

Mistake #4: Taking the correct particle test measurement

As-built cleanroom particle test is testing when cleanroom has just been completed. It typically does not have equipment or personnel in the room. At-rest cleanroom particle test is testing when the cleanroom has been in use and has equipment but no personnel are present and no operations are on-going. Operational cleanroom particle test is testing when personnel are in the cleanroom working and operating equipment. Your company QA / ISO / CGMP procedure will dictate which cleanroom particle test is required.

Mistake #5: Assuming continuous measurement particle measurement systems eliminate need for annual particle test

Some cleanrooms have continuous measurement particle measurement systems. These systems only take samples at a few fixed locations in each room. To meet  ISO-14644 and do an effective cleanroom particle testing requires more test locations than the fixed locations. So manual cleanroom particle tests should still be done on a regular basis to ensure your cleanroom is meeting its designed cleanroom classification.

Summary:

Regular and effective cleanroom particle testing is critical to keep your cleanroom operating properly. ISO14644-1 and 2 are the ISO standards for cleanroom classification and cleanroom particle testing. The 5 most common mistakes made in cleanroom particle testing are: 1) not enough sample locations 2) not using the correct particle counter 3) not having your particle counter calibrated regularly 4) not selecting the requirement cleanroom particle test and 5) relying on continuous measurement systems instead of regular cleanroom particle tests.

Property American Cleanroom Systems 2024

The author Anthony Chien has worked at American Cleanroom Systems for past 12 years. He has more than 40 years of cleanroom experience. Anthony has a BS and MS in Electrical Engineering specializing in semiconductor manufacturing from the University of Illinois. American Cleanroom Systems is a design build modular cleanroom manufacturer based in Rancho Santa Margarita CA. It is expert in ISO-5 through ISO-8 (class 100 to class 100k) modular cleanrooms for pharmaceutical cleanrooms, medical device cleanrooms, and industrial cleanrooms.

More Blog Posts

Read More
American Cleanroom Systems

Need a New Cleanroom Now?

Quotes in 48 hours. Call or Complete Online RFQ Form.

Request A Quote

Contact Us

23221 Arroyo Vista Rancho Santa Margarita California, 92688

Email: Info@AmericanCleanrooms.com

American Cleanroom Systems® 2024 - Unauthorized use and/or duplication of this material without express and written permission from this website owner is strictly prohibited. Excerpts and links may be used, provided that full and clear credit is given to American Cleanrooms with appropriate and specific direction to the original content.

Privacy Policy
 | 
Terms and Conditions
linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram